For endoscope washer disinfectors, endoscope storage/drying cabinets and instrument washer disinfectors, we provide all tests required periodically by HTM, WHTM, NI/HTM, SHTM and EN.
* It is critical that the time between sample taken, and sample testing is not greater than 24 hours.
Scope of Tests | Limit of Detection | Stability Time |
Total Viable Count (TVC) | 1cfu/100ml | 24 hours* |
Environmental Mycobacterium | 1cfu/100ml | 24 hours* |
Bacterial Endotoxins | 0.05EU/ml | 24 hours* |
Total Organic Carbon (TOC) | 0.09mg/L | 7 days |
pH | N/A | 5 days |
Electrical Conductivity | 0.15μS/cm | 1 day |
Total Dissolved Solids (TDS) | 8.23mg/L | 7 days |
Total Hardness | 1.11mg/L | 1 month |
Chloride | 0.256mg/L | 1 month |
Phosphate | 0.014 mg/L | 1 month |
Silicate | 0.030mg/L | 1 month |
Heavy Metals (Lead) | <0.1 mg/l | 1 month |
Iron | <0.0023 | 1 month |
Appearance | N/A | 1 month |